In Vivo. 2021 Nov-Dec;35(6):3431-3439. doi: 10.21873/invivo.12643.
ABSTRACT
BACKGROUND/AIM: Lack of interchangeability between prostate-specific antigen (PSA) assays could have a clinical impact. We compared PSA assays from different manufacturers and calibrations.
PATIENTS AND METHODS: A total of 233 men who underwent prostate biopsy (PSA: 2-10 ng/ml; Beckman Coulter Access® Hybritech® as reference) were enrolled. Total (tPSA) and free PSA (fPSA) were also measured using the Roche cobas® and the Abbott Architect® methods.
RESULTS: Roche tPSA values were ≈1% higher than Beckman, while Abbott values were ≈5% lower. Roche had the highest diagnostic sensitivity (92%) compared to Beckman Coulter (87%) and Abbott (85%). Roche fPSA was ≈3% lower and Abbott ≈17% higher than that of Beckman. For the percentage of fPSA, Roche had the highest sensitivity (98%).
CONCLUSION: Roche cobas® and Beckman Coulter Access® Hybritech® tPSA were almost interchangeable. While the agreement was acceptable for tPSA, this did not happen with fPSA and greater efforts for harmonization are required.
PMID:34697179 | DOI:10.21873/invivo.12643
Δεν υπάρχουν σχόλια:
Δημοσίευση σχολίου
Σημείωση: Μόνο ένα μέλος αυτού του ιστολογίου μπορεί να αναρτήσει σχόλιο.