Abstract
Purpose
With the increasing use of new minimally invasive approaches in temporal bone surgery, the need arises for evaluation of the risk of injury to sensitive anatomical structures. The factors that influence the measurement uncertainty (variation in representation of position and shape of anatomical structures) of imaging are of relevance. We investigate the effect of patients' anatomy on the measurement uncertainty of medical CT.
Methods
Six formalin-fixed temporal bones were used, fiducial markers were bone-implanted, and 20 CT scans of each temporal bone were generated. Surgically threatened anatomical structures of importance were defined. Manual segmentation was performed to create 3D surface models, and different Gaussian filters were applied. Analysis points were established along the border of the superior semicircular canal to determine the deviation between the 3D images of the labyrinth. The standard uncertainty was calculated, and one-way analysis of variance was performed (significance level = 5%) to evaluate the effect of certain factors (patient, side, Gaussian filter) on the measurement uncertainty.
Results
The influence of patient-specific anatomy on the measurement uncertainty of medical CT (p = 0.049) was demonstrated for the first time. The applied Gaussian filter (p = 0.622) and the patient's side (p = 0.341) showed no significant effect.
Conclusion
The applied method and the results of the statistical analysis suggest that the patient's individual anatomical conditions affect the measurement uncertainty of medical CT. Thus, the patient's anatomy must be considered as an important influencing factor during risk evaluation concerning minimally invasive and image-guided surgery.
https://ift.tt/2zWgcLu
Δεν υπάρχουν σχόλια:
Δημοσίευση σχολίου
Σημείωση: Μόνο ένα μέλος αυτού του ιστολογίου μπορεί να αναρτήσει σχόλιο.