Accuracy and reliability of a hand-held in vivo skin indentation device to assess skin elasticity

Abstract

Objective

The aim of this study was to evaluate the performance of a hand-held indentation device for fast and reliable determination of skin stiffness.

Methods

Device accuracy to indentation depths of 0.6 and 1.3 mm was first evaluated on plastic foam materials with mechanical properties verified by a laboratory material testing device. Subsequently, the device's sensitivity to detect age related changes in skin stiffness was evaluated among 46 healthy women (18-79 years). Finally, the reproducibility of the method was tested with 6 healthy subjects.

Results

High correlation was detected between indentation stiffness of reference material and Young's modulus determined with mechanical testing device (0.6 mm indenter: r=0.97, p=0.05; 1.3 mm indenter: r=0.98, p=0.04). Age-related decrease of 38% in skin stiffness was observed in healthy volunteers (p<0.05). The coefficient of variation for 0.6 and 1.3 mm indenters were 7.4% and 8.5%, respectively. No trend related to hysteresis effect was observed from repeated measurements.

Conclusions

The presented indentation technique was accurate against the laboratory material testing device. Furthermore, skin changes related to aging could be detected with the indentation technique. The new device was found to be feasible for monitoring skin stiffness in cosmetics and clinical conditions.