Abstract

Purpose

We present a new method to evaluate the accuracy of an eye tracker-based eye localization system. Measuring the accuracy of an eye tracker's primary intention, the estimated point of gaze, is usually done with volunteers and a set of fixation points used as ground truth. However, verifying the accuracy of the location estimate of a volunteer's eye center in 3D space is not easily possible. This is because the eye center, the center of corneal curvature, is an intangible point.

Methods

We evaluate the eye location accuracy by using an eye phantom instead of eyes of volunteers. For this, we developed a testing stage with a realistic artificial eye and a corresponding kinematic model, which we trained with \(\mu \text {CT}\) data. This enables us to precisely evaluate the eye location estimate of an eye tracker.

Results

We show that the proposed testing stage with the corresponding kinematic model is suitable for such a validation. Further, we evaluate a particular eye tracker-based navigation system and show that this system is able to successfully determine the eye center with a mean accuracy of 0.68 mm.

Conclusion

We show the suitability of the evaluated eye tracker for eye interventions, using the proposed testing stage and the corresponding kinematic model. The results further enable specific enhancements of the navigation system to potentially get even better results.