Αναζήτηση αυτού του ιστολογίου

Παρασκευή 10 Νοεμβρίου 2017

A Novel Method to Measure Artificial Eye Motility

imagePurpose: To measure objectively, reproducibly, and noninvasively artificial eye motility, a fundamental aspect in the (cosmetic) outcome of enucleation. Method: A gaze- and pupil-tracking system, the "iView X," was implemented to measure horizontal eye motility. The system, with adjusted software, was tested with patients wearing 1 prosthetic eye after enucleation for retinoblastoma. Measurements were repeated 5 times in every patient and analyses were performed twice by 2 independent observers. Reproducibility was tested via linear mixed models. After the implementation of the method, more data were obtained, including more patients with a history of enucleation for the treatment of retinoblastoma for eye tracker measurement, and differences in motility percentages between implant types and sizes were analyzed via Mann-Whitney U tests. Results: The intraclass correlation coefficient of the interobserver variable in the patient test group (n = 27, 6–53 years) was 0.98 and 0.96 for measurement of left gaze and right gaze, respectively. Intraobserver variation was